TP M 255 S Temperature Calibrator

The SIKA TP M 255 S solid temperature calibrators can be used everywhere, including on-site in workshops, test and measurement rooms as well as laboratories. The SIKA TP solid series covers a broad spectrum of industries with diverse applications.

The optimum thermal coupling from the block to test item is achieved with the correct adapter sleeve. Ideally, the internal diameter of the sleeve is 0.5 mm larger than the outer diameter of the test item. With the aid of the adapter sleeve, straight temperature sensors of virtually any length and width can be calibrated. The dry block covers the entire temperature range without the need to change the calibration medium. Viscosity, flash points or outgassing are of no concern.

Micro bath function

The use of calibration liquids offers certain advantages if temperature sensors with an unusual shape and size are to be tested. The test item is immersed directly into the liquid without an insulating air gap, resulting in direct temperature contact between the calibrator and the test item. The liquid, such as silicone oil, is chosen depending on the calibration temperature required. The continuous adjustment of the magnetic stirrer together with the removable sensor basket agitates the calibration liquid to create a large measuring zone. Furthermore, the sensor basket guarantees unhindered stirring and helps protect the tank floor.

Black body function

A patented infrared calibration sleeve is used to calibrate IR pyrometers or thermal imaging cameras. The special surface structure and the asymmetrical shapes create a „cavity radiator“ with an emission factor of 0.9994, prevent the reflection of interference radiation and emit the required temperature in an ideal form. The pyrometer is simply held at the specified distance above the measurement opening, thereby forming the desired measurement area on the bottom for the calibration to be performed. A support base can be fitted directly on the unit.

Features:

Dry block

Temperature range

Tolerance

Stability

 

RT...255 °C

±0,4 °C

±0,05 °C

Micro bath

Temperature range

Tolerance

Stability

 

RT...255 °C

±0,2 °C

±0,05 °C

Infrared

Temperature range

Tolerance

Stability

 

RT...255 °C

±0,5 °C

±0,05 °C

Resolution

0.01 °C in the range of RT...99.99 °C

else 0,1 °C

BlockØ 60 mm / depth 170 mm

Customer Service